Troubleshooting

日本語

Cannot connect

First check whether the physical device appears in the manufacturer’s PIXet Pro GUI. Then inspect USB, drivers, SDK/Python architecture, and the hard-coded SDK directory. Because connect() selects the first enumerated device, verify its logged identity when multiple devices are connected.

No physical detector found / FileDevice 0

The controller rejects a file device. Confirm physical detection in the GUI; never identify a file device as D08-W0313.

setThreshold, setBias, or setTimepixClock fails

Inspect the return code rc and lastError(). Do not confuse a THL DAC code with the API’s keV threshold. Check device-specific safe conditions before changing bias.

Acquisition fails

Inspect state() and its error string. Check TIFF/HDF5 output permissions, free space, data volume, and exceptions in callbacks. Acquisition is synchronous, so consider the context used to request a stop.

Unexpected pixel values

Check the Counting, Energy/ToT, or Time/ToA mode, threshold, and calibration. Do not treat uncalibrated ToT as energy.

Sources: supplied pixet_ctrl.py, MiniPIX Datasheet, PIXet Pro User’s Manual.