D08-W0313 individual calibration

日本語

This page records the individual D08-W0313 calibration protocol dated 2025-02-11. Never reuse these factory calibration values for another detector. The listed PIXet version, 1.7.8.936, is the calibration-time version and may differ from the runtime SDK.

Item

Certificate value

Timepix clock

10.0 MHz

Ikrum

1

FBK

156

THL Coarse

8

Preamp

225

Equalization mean THL

751.666

Equalization standard deviation

2.250

Equalization masked pixels

2

Final THL

707

Threshold calibration A / B

11.831 / 49.262

Minimum threshold

3.697 keV

THL at 5 keV

691.585 (summary: THL=691)

Bad pixels under X-ray test

83

Largest bad-pixel cluster

66

Ten-minute stability

99.469%

ToT energy resolution σ near 60 keV

2.728 keV

The digital/bias test found zero bad pixels, while the X-ray test found 83: these are different tests, not conflicting measurements of one metric. Bias leakage was tested from 15 to 150 V; the tabulated leakage current is 0.0000 µA at each point, which does not establish a universal zero-leakage guarantee.

The mean ToT energy calibration coefficients are A=1.642, B=22.945, C=19.630, and T=4.368. These are certificate calibration parameters, not values to pass directly to set_threshold(). The Python API’s set_threshold(threshold_kev) requests a threshold in keV through PIXet.

Both imaging (Counting) and tracking (Energy) are classified B / Usable for this unit; this does not guarantee performance under every experimental condition.

Source: D08-W0313 Calibration Protocol pp. 1–4.